Common faults and solutions of Techcon TS250/TS350 dispensing machine

Release Date:2017-07-17

Common faults and solutions of Techcon TS250/TS350 dispensing machine  

fault

Possible reasons

Troubleshooting methods

LCDNot lit up

• Not powered on

Check the power cord connection
• Turn on the power

The system cannot start

• The 'foot pedal' switch is not connected or the plug is not inserted correctly
There is an issue with the foot switch
The internal wiring of the device is damaged or the connection is not secure
There is a problem with the solenoid coil
PCThere is a problem with the board

Check the connection of the foot switch
• Repair or replace the foot switch
Unplug the power cord and disconnect the air source. Remove the cover and check for damaged wires or loose connections
• Replace the solenoid coil
• ReplacePCplate

The system cannot boost the pressure

• Insufficient air pressure
The air hose is not connected
There is a problem with the regulator

Increase the pressure of the gas source
• Check the connection
• Replace the regulator

The system cannot be evacuated

The vacuum setting is too low
There is a problem with the solenoid coil
There is an issue with the vacuum Venturi tube
There is a problem with the vacuum needle valve

• Increase the vacuum setting
Replace the solenoid valve
• Replace the Venturi tube
Replace the needle valve

Inconsistent glue dispensing

There are bubbles in the adhesive
The dispensing time is too short
At the beginning, the dispensing needle is blocked

• Reduce the vacuum setting
• Extend the dispensing time
• Replace the needle



For more information about Techcon TS250/TS350 dispensing machines, please feel free to consult with Techcon's high-quality agent, Shenzhen Yapude Technology Co., Ltd //www.kitechain.com